A hospital is a health care institution providing patient treatment by specialized staff and equipment. In accord with the original meaning of the word, hospitals were originally "places of hospitality", and this meaning is still preserved in the names of some institutions such as the Royal Hospital Chelsea.

There are over 17,000 hospitals in the world.


Search This Blog

June 14, 2012

Sterilization Of laparoscopy Instruments

How laparoscopy Instruments sterilized:

After finished the procedure the instrument should clean properly.A method of cleaning that is ultrasonic cleaning . This method is the most efficient and effective. In fact, ultrasonic cleaning is 16 times more efficient than hand-cleaning. The instruments are placed in the ultrasonic unit for 10-15 minutes and use a neutral pH solution. Attention should be given to the following points during ultrasonic cleaning:

  • Before placing into the ultrasonic unit, the instruments are cleaned of all visible debris.
  • It’s preferable not to mix instruments made of dissimilar metals (such as aluminum and stainless) in the same cycle.
  • As with all types of cleaning, all instruments should be opened so ratchets and box locks are fully exposed to the cleaning process.
  • Upon completion of the cycle, the instruments are removed immediately and rinsed.
After cleaning The instruments should be dried and rinsing cycle before they are packed for sterilization. 

ETO
This is ideally achieved by using an air gun that blows all the water droplets off the surfaces of instruments. . Then the instruments packaged and submitted to sterilization by ethylene oxide .Using ethylene oxide (EO) would work for all disposable instruments, insulated hand instruments and tubings employed for gas, suction and irrigation. Endoscopic instruments may be sterilized with either cold or warm EO gas, with respect to the manufacturer’s instructions. With cold gas, the temperatures are set at 85 0C and also the instruments are subjected for 4 hours and 30 minutes. Aeration must then follow for 12 hours. Warm gas sterilization happens at 145 0C for 2 hour 30 minutes, followed by 8 hours aeration. The benefits of EO are the items aren't damaged, it's non-corrosive to optics also it permeates porous material. Its main disadvantages are its cost, toxicity, the requirement for aeration and being a longer process. 

The laparoscopy  instrument may be sterilized by 2% glutaraldehyde, 6% stabilized hydrogen peroxide and per acetic acid (acetic acid/hydrogen peroxide). Fibreoptic light cords and telescopes have to be soaked in 2% glutaraldehyde not less than Ten minutes. Soaking should not exceed Twenty minutes. The endocamera could also disinfected by 10 minutes submersion in 2% glutaraldehyde. Care must be come to leave the plug end of the cord away from solution. Alternately, sterile drape over the camera and cord may be used. Soakage of other metallic instruments, including trocars, and hand instruments, has become recommended for An hour, to avoid infection with atypical mycobacterial infection. Formaldehyde, glutaraldehyde from phenolic derivatives, iodophors, hypochlorites, phenolics and quartery ammonium compounds are unpopular and it has been condemned. Formaldehyde is potentially cancer causing and very irritating to the skin, eyes, nose, and respiratory tract. Furthermore, its efficacy is found wanting, and for that reason, routine utilization of formaldehyde for sterilizing instruments and other items isn't recommended. 

 STERRAD 
 One of the newer sterilizer system - STERRAD (Johnson & Johnson) - uses hydrogen peroxide vapor and low-temperature gas plasma to sterilize most devices quickly with no toxic residues. Usually, the process takes about 75 minutes for wrapped and dry instruments and devices. Within the chamber, a deep vacuum is drawn. Fifty-nine percent aqueous peroxide is vaporized into the chamber. The product will be enveloped within the peroxide vapor. Following a diffusion of the gaseous peroxide with the load, chamber pressure is reduced, permitting the generation of low-temperature gas plasma. Rf (RF) energy is put on the chamber via an RF amplifier, inducing the plasma state. Reactive species are generated in the peroxide on this state, reacting with materials and every other. When the high-energy species have reacted, they recombine to form water vapor, oxygen, along with other non-toxic by products. Upon completion of sterilization, instruments are dry for immediate use or sterile storage. Thus, recontamination risk is minimized, and given that they remain sterile until their next use, money and time is saved by avoiding reprocessing instruments when the case in canceled or delayed. This system occupies minimal space and requires no venting or water hookup. The only utility requirement is electrical hookup. 